ORP TRANSMITTER T80 – ECD
Product Type: ORP TRANSMITTER T80
Model: T80-10-210-01
Manufacturer: ECD (USA)
Application:
Used for monitoring ORP parameters from ORP Sensor.
Main Display screens; the Data Screen, the Millivolt Screen and the Graphical Display screen.
The Data Screen displays the measurement type, the measured value with units, the % milliamp output of the 4-20 mA channel and the temperature. The mV Screen displays the measurement type, the raw millivolt signal from the sensor, the % milliamp output of the 4-20 mA channel and the temperature. The Graphical Screens display the measurement type, the measured value with units and a graphical representation of the % milliamp output. Three graphical styles are available; a Trend line, a Bar graph or a Gauge. The status of alarm relays, energized/de-energized is displayed on transmitters with relays.
Product Type: ORP TRANSMITTER T80
Model: T80-10-210-01
Manufacturer: ECD (USA)
Application:
Used for monitoring ORP parameters from ORP Sensor.
Main Display screens; the Data Screen, the Millivolt Screen and the Graphical Display screen.
The Data Screen displays the measurement type, the measured value with units, the % milliamp output of the 4-20 mA channel and the temperature. The mV Screen displays the measurement type, the raw millivolt signal from the sensor, the % milliamp output of the 4-20 mA channel and the temperature. The Graphical Screens display the measurement type, the measured value with units and a graphical representation of the % milliamp output. Three graphical styles are available; a Trend line, a Bar graph or a Gauge. The status of alarm relays, energized/de-energized is displayed on transmitters with relays.
Specifications:
- Have 2 types: single or dual channel.
- Input connection: 1 or 2 ORP Sensor (or another type sensor).
- Input range: -1500 – +1500 mV.
- Temperature: -30 – 140°C.
- Accuracy: ±1 mV
- Enclosure: Polycarbonate, NEMA 4X, weatherproof, ½ DIN, (L xWx D) 5.7” X 5.7” X 3.5” (14.4cm X 14.4cm X 9.0cm).
- Display: 128 x 64 pixels (2.75” x 1.5”) LCD, Black/Grey background on loop powered instruments, Blue/White background LED backlight on 100-240 VAC and 24 VDC.
- Input power:
Code -0: Loop powered, 24 VDC, 600 Ω maximum load (18-36VDC @ 35 mW minimum)
Code -1: 24 VDC (18-36 VDC @ 250 mW minimum)
Code -2: 100-240 VAC, 50/60 Hz, 4W.
- Outputs: 4-20 mA output (standard), Fault Condi0on: 3.5 mA, 22 mA or none, Modbus RTU (standard); HART® (optional); Alarm Relays (Op0onal) Three (3) SPDT, form 1C, 250 VAC, 3 Amp resistive maximum relays, user configurable as Hi/Lo or Fault alarms.
Warranty policy from HABCOM Technology Corporation
HABCOM warrants all products following manufacturer’s standards.
All warranted products must be used under manufacturer’s conditions.
Supporting time: Within working hours (8h-12h & 13h-17h) on working days of the week.
For important devices – Our company will have spare parts to replace as the commitments.